Chapter 7: Printer
Pulse Oximeter for Sleep Screening Operation Manual 7-1
Chapter 7: Printer
Description
Up to 12 hours of previously stored data can be printed as a summation of data from the
monitors built-in printer or as a data download to a computer.
Trend Summary Printout
To print the trend summary from the monitor’s built-in printer:
1. The monitor must be in the Trend Printing Mode.
2. The Trend Printing Mode will be entered after the monitor is powered down in the Data
Collection Mode or when the trend memory becomes full.
3. Press the
F key to print out the summary report.
4. Press the F key again to stop printing or unit will stop printing after the summary report
has been printed.
The trend summary printout from the monitor’s built-in printer will contain the following
sections: (see Figure 7.1: Sample Trend Printout)
Report Title – The first section will contain a report title to indicate that a new printout
has begun.
Patient Information – The patient information section supplies a space for the user to write
the following patient information on the printout:
Patient name
Date
Diagnosis
Comments
Start time
Study Duration – The study duration section reports the total time consumed by separating
the data into the following three items:
Total duration
Excluded time (due to invalid data caused by sensor disconnect, artifact, etc.)
Total valid sample
Study Maximums and Minimums – The study maximums and minimums section contains
the following information:
Low SpO
2
– The lowest recorded SpO
2
value taken from the total valid sample.
Time of Lowest SpO
2
– Records the time the lowest SpO
2
reading occurred (clock
does not ignore invalid data when determining lowest SpO
2
time).
High PR – The highest recorded pulse rate value taken from the total valid sample.
Low PR – The lowest recorded pulse rate value taken from the total valid sample.
Avg. SpO
2
– The average SpO
2
value from the total valid sample.